B&K Precision 866C Uživatelský manuál Strana 69

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 76
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 68
The stored calibration data will be used the next time the DC Load is power cycled.
0x67 Set calibration information
Byte
offset
Meaning
3 to 22 ASCII information representing the calibration. Example: you might wish to
store the date and time of the calibration and the initials of the person who
performed the calibration.
23-24 Reserved
0x68 Read calibration information
Byte
offset
Meaning
3 to 22 ASCII information representing the calibration.
23-24 Reserved
0x69 Restore the factory default calibration data
Byte
offset
Meaning
3-24 Reserved
0x6A Get product's model, serial number, and firmware version
Byte
offset
Meaning
3 to 7 ASCII model information
8 Low byte of firmware version number
9 High byte of firmware version number
10 to 19 Instrument's serial number in ASCII
20-24 Reserved
0x6B Read the bar code information
Note: the bar code information is to be interpreted as ASCII data.
Byte
offset
Meaning
3 to 5 Identity
6 to 7 Sub
7 to 9 Version
10 to 11 Year
8500 DC Load Series Version: November 4, 2008 Page 69 of 76
Zobrazit stránku 68
1 2 ... 64 65 66 67 68 69 70 71 72 73 74 75 76

Komentáře k této Příručce

Žádné komentáře